Skip to main content
. Author manuscript; available in PMC: 2022 Oct 1.
Published in final edited form as: Curr Protoc. 2021 Oct;1(10):e279. doi: 10.1002/cpz1.279

Table 5:

Troubleshooting

Problem Possible Cause Solution
Inflated lungs during perfusion, liquid coming from mouth or nose The needle might have inserted into the right atrium or right ventricle instead of the left ventricle Remove the needle and insert it into the left ventricle
Fixation tremor is not observed during paraformaldehyde perfusion/fixation The needle might have inserted into the wrong place Re-insert the needle into the left ventricle
Blood vessels can be seen in the perfused brain Perfusion was unsuccessful Perform perfusion by inserting a syringe needle in the left ventricle and make a cut in the right atrium

Increase the volume of solutions used or the timing of the perfusion process
The brain is too soft in clearing solution The brain was not fixed properly Perform post-extraction fixation for a longer time 48-72 hours in 4 % paraformaldehyde (in PBS)
Brain tissue becomes unclear/opaque after storing in PBS for a long time The tissue might have absorbed salt Perform TDE clearing for 24-48 hours
Fluorescence is too low Antibody concentration might not have been enough,

Exposure duration was not enough
Use a higher concentration of antibody and incubate for a longer period
Vasculature/blood vessels are observed during astrocyte imaging Perfusion was unsuccessful Perform perfusion by inserting a syringe needle in the left ventricle and make a cut in the right atrium

Perform perfusion with a higher volume of PBS
The fluorescence signal is not obtained from the inner part of the tissue Antibody might have difficulty in reaching inner sections of the brain Perform longer incubation and use 1:50-1:100 concentration of antibodies

Use polyclonal booster antibodies (such as anti-GFP alexa 488 which will enhance GFP fluorescence at the same fluorescence channel)

If possible, perform brain sectioning (hemibrains) or slicing so that antibodies can reach to inner parts easily