Skip to main content
. 1999 Oct;37(10):3102–3107. doi: 10.1128/jcm.37.10.3102-3107.1999

TABLE 3.

Assessment of LCx assay results after analysis for resolution of discrepant results (culture result plus clinical diagnosis)

Specimen (no.) No. of specimens MTBC positive by culture
No. of specimens MTBC negative by culture
Sensitivity (%) Specificity (%) PPV (%) NPV (%)
LCx assay positive LCx assay negative LCx positive LCx assay negative
All (2,347) 115 50 1 2,139 69.7 99.9 99.1 97.7
Respiratory (2,083) 92 39 1 1,915 70.2 99.9 98.9 98.0
 Smear positive (139) 65 1 0 60 98.5 100 100 98.4
 Smear negative (1,944) 27 38 1 1,855 41.5 99.9 96.4 98.0
Nonrespiratory (264) 23 11 0 224 67.6 100 100 95.3