Table 2-.

Results of femoroplasty: YL and ML represent the yield load and maximum load in N, respectively, and YE and ME represent the yield energy and maximum energy in Joules, respectively. Moreover, “Pre”, “Plan”, and “Post” represent the FE analysis based on the pre-operative CT, virtual CT created from the plan, and post-operative CT, respectively.

Control Augmented
Simulation Fracture test Simulation Fracture test
Sample # YL [N] YL [N] ML [N] YE [J] ME [J] YL [N] (Pre) YL [N] (Plan) YL [N] (Post) YL [N] ML [N] YE [J] ME [J]
1 1500 1419 1706 2.90 20.07 1350 2785 1910 2282 2360 8.91 19.03
2 1825 2302 2497 3.77 26.97 1745 3330 2800 3003 3142 8.96 58.26
4 1815 1771 2253 4.22 28.72 1680 2830 2675 2685 2685 4.50 13.33
5 1725 2145 2145 4.90 19.08 1170 2395 2940 3087 3159 22.67 39.63
6 2465 2116 2481 6.37 25.17 2350 3045 3040 2812 2920 7.48 24.19
Mean 1866 1951 2216 4.43 24.00 1659 2877 2673 2774 2853 10.50 30.89
SD 359 355 322 1.3 4.2 453 345 448 317 336 7.0 18.2