Table 3:

Measured concentrations for iodine and hydroxyapatite compared with ground truth.

Iodine Hydroxyapatite
True conc. (mg/cc) Measured conc. Mean ± SD (mg/cc) Difference True conc. (mg/cc) Measured conc. Mean ± SD (mg/cc) Difference
2.000 2.137 ± 0.26 0.137 200.000 197.090 ± 4.94 −2.91
5.000 5.004 ± 0.25 0.004 400.000 380.062 ± 10.59 −19.94
10.000 9.998 ± 0.33 −0.002
15.000 14.345 ± 0.38 −0.655
RMSE = 0.33 mg/cc RMSE = 14.24 mg/cc