Table 1.

Summary of the scan configurations used for the image acquisitions described below.

TCM calibration/simulator evaluation TCM module evaluation Pilot VIT

Scanner Force, DukeSim DukeSim DukeSim
Phantom Mercury ATOM (adult, male) XCAT (BMIs of 24.3, 28.2, and 33.0)
graphic file with name nihms-1747253-t0013.jpg graphic file with name nihms-1747253-t0014.jpg graphic file with name nihms-1747253-t0015.jpg
Source of TCM profiles CD4D CD4D, DukeSim DukeSim
mAsref 50, 100, 200 100 100
Average CTDIvol (mGy per mAs) 2.36, 4.52, 8.92 (for each mAsref) 3.77, 3.98 (for each TCM profile) 7.27, 10.01, 14.02 (for each phantom)
TCM strength Very weak, weak, average, strong, very strong Average Very weak, weak, average, strong, very strong
Lref (cm) 33.9 33.9 31.4
Helical pitch 0.6 0.6 1.0