Table 1. Performance Comparison of Existing State-of-the-Art Cortisol Aptasensors.

method probe surface test matrix DR (nM) LOD (nM)
colorimetric aggregation of AuNP37 aptamer AuNP serum 150–600 29.5
electrochemistry (amperometry) triamcinolone competitive assay38 aptamer graphene-modified glassy carbon electrode saliva and serum 0.1–32 000 0.032
aptamer-based lateral flow assay40 aptamer–AuNP cysteamine/nitrocellulose sweat 32–327 2.7
aptamer-based lateral flow assay13 aptamer–AuNP nitrocellulose diluted saliva 0.41–41 1.01
copper porphyrin-printed electrode41 aptamer–magnetic particle multiwalled carbon nanotubes/copper/porphyrin saliva 1.6–65.0 1.0
CNT–nanogap FET39 aptamer multiwalled carbon nanotube serum (10× diluted) 0.1–2000 50
surface plasmon resonance (SPR)42 aptamer gold/titanium layer over glass sweat 82–16 500 32
electrochemistry (EIS)43 aptamer mercaptohexanol over gold electrode sweat 5–30 5
electrochemistry (DPV) antibody–AuNP prelabeled cortisol58 aptamer–antibody with AuNP MWCNTs/CMK-3/AgNP/AuNP saliva (10× diluted) 0.00002–276 3 × 10–7
electrochemistry (DPV) (this work) structure-switching aptamer–MB AuNW/BSA/GA over gold electrode serum 1–1000 0.68