Skip to main content
. 2021 Oct 14;6(42):28403–28409. doi: 10.1021/acsomega.1c04792

Table 1a. Results of the Accuracy Test with Three Different Concentrations of HPLC-DAD and the GC-FID Methoda (Revised).

  DNPH method HPLC-DAD
  p-toluenesulfonic acid liquid GC-FID
concentration (μg/mL) %recovery RSD (%) concentration (μg/mL) %recovery RSD (%)
1 101.16 0.22 400 101.50 1.48
2 101.15 0.50 500 101.37 0.56
3 100.82 0.59 600 101.99 1.02
a

Acceptance criteria: %Recovery: 95–105%; RSD: 3.7% DNPH (2,4-dinitrophenylhydrazine). HPLC-DAD (high-performance liquid chromatography-diode array detector). SHS-GC-MS (static headspace-gas chromatography-mass spectrometry). GC-FID (gas chromatography-flame ionization detection). RSD (relative standard deviation).