Table 11.
Spatial-peak temporal-average intensity of diagnosis acoustic transfer for nonfetal Doppler application[441].
| Location/use | Intensity [mW cm−2] |
|---|---|
| Peripheral vessel | 720 |
| Cardiac system | 430 |
| Fetal imaging and other | 94 |
| Ophthalmic sites | 17 |
Spatial-peak temporal-average intensity of diagnosis acoustic transfer for nonfetal Doppler application[441].
| Location/use | Intensity [mW cm−2] |
|---|---|
| Peripheral vessel | 720 |
| Cardiac system | 430 |
| Fetal imaging and other | 94 |
| Ophthalmic sites | 17 |