Table 2.

Power requirements of implantable and ingestible biomedical electronic devices.

Implantable/Ingestible device Power requirement References
Deep brain stimulator (DBS) 100 μW [104,105]
Spinal cord stimulator 1–10 mW [104]
Intraocular pressure (IOP) monitor 200 nW–200 μW [106]
Retinal prosthesis 250 mW [107]
Cochlear implant 100 μW–10 mW [105,108,109]
Pacemaker 10–30 μW [79,102]
Implantable cardioverter defibrillator (ICD) 50–500 μW [105]
Implantable blood pressure monitor Passive
Wireless capsule endoscope 5–30 mW [110]
Medication adherence monitor 1 mW [35]
Gastric stimulator 1–30 mW [111]
Implantable drug delivery system 100 μW–1 mW [104,105,112]
Artificial urinary sphincter 200 μW [79,102]