. 2021 Oct 29;102(2):115587. doi: 10.1016/j.diagmicrobio.2021.115587

Table 3A.

Rates of positive results (sensitivity) of 5 commercial assays for SARS-CoV-2 antibodiesa. Anti–SARS-CoV-2 ELISA (IgG) EUROIMMUN and Anti–SARS-CoV-2 ELISA (IgA) EUROIMMUN were not considered separately. Rather, the combination of both tests was evaluated.

Assay/Db Nc SARS-CoV-2 IgG
Abbott Diagnostics
(Abbott-IgG)
Anti–SARS-CoV-2 Roche
(Roche-IgT)
Anti–SARS-CoV-2 ELISA (IgG) EUROIMMUN
(EUROIMMUN-IgG)
Anti–SARS-CoV-2 ELISA (IgA) EUROIMMUN
(EUROIMMUN-IgA)
Anti–SARS-CoV-2
ELISA (IgA) and (IgG)
EUROIMMUN
combinedc
(EUROIMMUN-IgA/G)
LIASON SARS-CoV-2
S1/S2 IgGd
(DiaSorin-IgG)
0 16 4 (25) 3 (19) 2 (13) 4 (25) 4 (25) 3 (19)
1 13 1 (8) 0 (0) 0 (0) 1 (8) 1 (8) 2 (15)
3 26 9 (35) 8 (31) 6 (23) 10 (38) 12 (46) 4 (15)
5 13 (12) 7 (54) 7 (54) 6 (46) 7 (54) 7 (54) 5 (42)
6 11 7 (64) 7 (64) 6 (55) 7 (64) 7 (64) 8 (73)
7 11 (10) 7 (64) 6 (55) 4 (36) 8 (73) 8 (73) 5 (50)
9 18 14 (78) 15 (83) 14 (78) 16 (89) 16 (89) 14 (78)
10 11 (10) 10 (91) 10 (91) 9 (82) 11 (100) 11 (100) 9 (90)
12 18 (17) 16 (94) 17 (100) 13 (76) 17 (100) 17 (100) 14 (88)
14 15 (14) 15 (100) 14 (93) 13 (87) 14 (93) 14 (93) 13 (93)
16 14 (13) 13 (93) 13 (93) 13 (93) 13 (93) 13 (93) 12 (92)
20 10 10 (100) 10 (100) 10 (100) 10 (100) 10 (100) 10 (100)
35 10 9 (90) 10 (100) 10 (100) 10 (100) 10 (100) 10 (100)
44 5 4 (80) 5 (100) 5 (100) 5 (100) 5 (100) 5 (100)
a

The cells contain absolute numbers and percentages in brackets.

b

If less than 10 samples were available for a specific day, samples from several days were combined. The displayed day number is then the mean for the group.

c

The Anti–SARS-CoV-2 ELISA (IgG) EUROIMMUN and Anti–SARS-CoV-2 ELISA (IgA) EUROIMMUN were not considered separately. Rather, the combined test was considered positive if any of the tests was positive.

d

There are 6 less samples for LIASON SARS-CoV-2 S1/S2 IgG, the number of samples for this test are in brackets in the second column.