Skip to main content
. 2021 Oct 30;145:105021. doi: 10.1016/j.jcv.2021.105021

Table 1.

Analytical performances of the COVID-19 ID NOW test assay performed in emergency department and ID Now test in microbiology laboratory compared to reference RT-PCR assay.

RT-PCR Tests performance (%)
Detected Not detected
ID Now test performed in ER
Detected
Not detected

151
3

6
235
Sensitivity: 98.0
Specificity: 97.5
PPV: 96.2
NPV: 98.7
ID Now test in microbiology laboratory
Detected
Not detected

100
60

0
235
Sensitivity: 62.5
Specificity: 100
PPV: 100
NPV: 79.7