TABLE 3.

Optimum cutoffs based on the ROC curves and their performancesa

Assay (n) Cutoff Sensitivity Specificity Overall agreement (%)
Genscript cPass (125) Manufacturer’s cutoff, 30% 71% 94.23% 81.6%
Based on ROC cutoff, 20% 92.75% 87.5% 90.4%
P value <0.001 0.065 <0.05
Diasorin-TrimericS IgG (103) Manufacturer’s cutoff, 33.8 BAU/ml 84.91% 90% 87.37%
Based on ROC cutoff, 40 BAU/ml 84.90% 96% 90.29%
P value 0.998 0.092 0.506
Roche Elecsys (96) Manufacturer’s cutoff, 0.8 U/ml 100% 41.67% 70.83%
Based on ROC cutoff, 5 U/ml 93.75% 70.83% 82.29%
P value <0.05 <0.001 0.0616
a

The other three tests showed ROC-based cutoffs similar to the manufacturer’s values.