TABLE 1.

Example of dual detector parameter set from a single patient and hemisphere.

Setting Value Notes
Stimulation rate 130.2 Hz Stimulation rate
Stimulation pulse 60 μs Stim pulse width
Stimulation contact +1-case Monopolar stimulation
Shared parameters for both detectors
Time domain sampling rate 500 Hz Sampling rate for time domain data
LPF1 450 Hz Embedded low pass filter before amplification
LPF2 1,700 Hz Embedded low pass filter after amplification
FFT interval 500 ms Interval in which FFT is computed
FFT size 1,024 Number of points for onboard FFT computation
Ramp up rate 0.03 mA/sec Rate at which stimulation changes from lower to higher amplitudes
Ramp down rate 0.12 mA/sec Rate at which stimulation changes from higher to lower amplitudes
Motor detector
Sense channel + 9−8 MC (motor cortex)
Linear detector power band input 64.45 – 66.41 Hz Predefined power band for embedded on board power computation
Update rate 60 Number of FFT intervals averaged in non-moving average, represents 30 s of data. Defines algorithm rate (e.g., algorithm state is determined every 30 s). This is referred to as “detector count” below.
Onset 0 Number of detector counts must be above threshold to change to state. A value of 0 means that as soon as the threshold is crossed stimulation ramps to the target state.
Termination 4 Number of detector counts must be below threshold to change to state. Detector value must be below threshold for 2 consecutive minutes (update rate × 4; = 4 min in this case) before it transitions to state.
State change blank 30 In units of FFT interval. On state change power values are not computed into linear detector for 15 s (FFT interval × 30).
Target current state 0 3.1 mA Target stim to ramp to when stimulation is below lower threshold (for the first linear detector- “LD0”)
Target current state 1 HOLD Target stim to ramp to when stimulation between upper and lower thresholds (for the first linear detector – “LD0”)
Target current state 2 2.7 mA Target stim to ramp to when stimulation is above lower threshold (for the first linear detector – “LD0”)
Sleep Detector:
Sense channel +9−8 MC (motor cortex)
Linear detector power band input 3.42 – 12.21 Hz Predefined power band for embedded on board power computation
Update rate 60 Number of FFT intervals averaged in non-moving average, represents 30 s of data. Defines algorithm rate (e.g., algorithm state is determined every 30 s)
Onset 0 Number of detector counts must be above threshold to change to state (in “update rate” units).
Termination 10 Number of detector counts must be below threshold to change to state. Detector value must be below threshold for 5 consecutive minutes (update rate × 10; = 10 min in this case) before it transitions to state.
State change blank 30 In units of FFT interval. On state change power values are not computed into linear detector for 15 s (FFT interval × 30).
Target current state 3 2.7 mA Target stim to ramp to when stimulation is below lower threshold (for the second linear detector – “LD1”)
Target current state 4 2.7 mA Target stim to ramp to when stimulation between upper and lower thresholds (for the second linear detector – “LD1”)
Target current state 5 2.7 mA Target stim to ramp to when stimulation is above lower threshold (for the second linear detector – “LD1”)