TABLE II. SUMMARY OF ACCELERATED AGING TEST FOR GRAPHENE MICROELECTRODE ARRAYS.
Array | Final Status | Time Elapsed (days in 87 °C PBS) | Equivalent Lifetime (years in 37 °C) | Failure Mechanism |
---|---|---|---|---|
PET #1 | Working | 30 | 2.63 | No failure |
PET #2 | Working | 30 | 2.63 | No failure |
PET #3 | Working | 30 | 2.63 | Minor PBS Permeation |
PET #4 | Working | 30 | 2.63 | No Failure |
SU-8 #1 | Failed | 22 | 1.93 | SU-8 cracks over electrode |
SU-8 #2 | Failed | 17 | 1.49 | SU-8 cracks over electrode |
SU-8 #3 | Failed | 8 | 0.70 | SU-8 cracks over wire |
SU-8 #4 | Failed | 7 | 0.61 | SU-8 cracks over wire |