Skip to main content
. Author manuscript; available in PMC: 2021 Nov 2.
Published in final edited form as: IEEE Trans Biomed Eng. 2020 Mar 17;67(11):3203–3210. doi: 10.1109/TBME.2020.2979475

TABLE II. SUMMARY OF ACCELERATED AGING TEST FOR GRAPHENE MICROELECTRODE ARRAYS.

Array Final Status Time Elapsed (days in 87 °C PBS) Equivalent Lifetime (years in 37 °C) Failure Mechanism
PET #1 Working 30 2.63 No failure
PET #2 Working 30 2.63 No failure
PET #3 Working 30 2.63 Minor PBS Permeation
PET #4 Working 30 2.63 No Failure
SU-8 #1 Failed 22 1.93 SU-8 cracks over electrode
SU-8 #2 Failed 17 1.49 SU-8 cracks over electrode
SU-8 #3 Failed 8 0.70 SU-8 cracks over wire
SU-8 #4 Failed 7 0.61 SU-8 cracks over wire