. 2021 May 21;4(3):143–148. doi: 10.1016/j.jimed.2021.05.005

Table 3.

Comparison of radiation dose and operation time among the three groups.

Dose parameters Conventional group optimized length group Low-dose optimized length group Con/ Opti
Opti/Low
P P
positioning DLP(mGy·cm) 312.4 ​± ​143.9 287.5 ​± ​130.1 33.87 ​± ​11.99 0.274 <0.001
guided DLP(mGy·cm) 294.3 ​± ​259.4 185.3 ​± ​157.4 21.34 ​± ​15.10 0.003 <0.001
review DLP(mGy·cm) 241.7 ​± ​197.2 220.2 ​± ​141.3 27.10 ​± ​14.55 0.451 <0.001
Total mAs 6211 ​± ​3043 5122 ​± ​2111 1070.6 ​± ​306.7 0.013 <0.001
Average CTDIvol(mGy) 9.79 ​± ​3.75 9.44 ​± ​3.04 1.33 ​± ​0.38 0.531 <0.001
ED(mSv) 11.95 ​± ​5.86 9.77 ​± ​4.07 1.23 ​± ​0.41 0.010 <0.001
Operation duration(min) 25.85 ​± ​8.95 22.40 ​± ​7.45 21.42 ​± ​7.91 0.021 0.214