Table 3. OCT/OFDI analysis at stent edges.

Overall Edge with dissection Edge without dissection p-value
(N = 345) (N = 35) (N = 310)
    Reference lumen area, mm2 5.3 (4.0–7.3) 4.3 (3.8–5.7) 5.4 (4.1–7.4) 0.0009
    Stent area at stent border, mm2 6.1 (4.7–7.8) 6.0 (4.7–7.6) 6.1 (4.6–7.8) 0.82
    Lumen area at adjacent site of stent edge, mm2 5.7 (4.3–7.7) 5.0 (4.0–6.6) 5.7 (4.3–7.7) 0.07
    Stent-oversizing index 1.1 (1.0–1.3) 1.4 (1.1–1.7) 1.1 (1.0–1.3) <0.0001
    Lumen long diameter / lumen short diameter 1.1 (1.1–1.2) 1.2 (1.1–1.3) 1.1 (1.1–1.2) 0.009
    Lumen eccentricity 0.12 (0.09–0.17) 0.16 (0.12–0.24) 0.12 (0.09–0.16) 0.002
    Type of plaque at stent edge 0.87
    Normal 30 (8.7) 0 (0.0) 30 (9.7)
    Fibrous 208 (60.3) 8 (22.9) 200 (64.5)
    Lipidic 64 (18.6) 16 (45.7) 48 (15.5)
    Fibrocalcific 43 (12.5) 11 (31.4) 32 (10.3)
Progressive atherosclerotic plaque 107 (31.0) 27 (77.1) 80 (25.8) <0.0001
Location of dissection
    Proximal - 12 (34.3) -
    Distal - 23 (65.7) -
Length of flap, mm - 0.9 (0.4–1.4) -
Thickness of flap, mm - 0.3 (0.2–0.4) -
Length of dissection, mm - 2.4 (1.5–3.4) -
Arc of dissection, ° - 45 (32–82) -
Depth of cavity, mm - 0.3 (0.2–0.4) -
Functional lumen area, mm2 - 3.7 (2.6–5.6) -
Depth of dissection
    Intima - 22 (62.9) -
    Media - 13 (37.1) -
    Adventitia - 0 (0.0) -
Hematoma - 2 (5.7) -

Values are presented as median (interquartile range), or n (%) for categorical variables. Generalized estimating equation (GEE) method with gamma with log-link model and was used for continuous valuables and GEE with ordinal logistic model was used for categorical data. OCT, optical coherence tomography; OFDI, optical frequency domain imaging.