Table 4.

Validation results for the three devices

Part 1 ≤5 mmHg ≤10 mmHg ≤15 mmHg Grade 1 Mean (mmHg) SD (mmHg)
Pass required
 Two of 73 87 96
 All of 65 81 93
Omron HEM-7120 achieved
 SBP 73 94 96 Pass 1.3 5.3
 DBP 73 91 96 Pass -2.0 5.6
Yuwell YE680A achieved
 SBP 50 79 91 Fail 1.0 9.0
 DBP 67 87 95 Fail -1.1 6.5
Cofoe KF-65B achieved
 SBP 53 79 91 Fail 4.1 7.0
 DBP 65 89 96 Pass -3.3 5.3
Part 2 2/3 ≤5 mmHg 0/3 ≤5 mmHg Grade 2 Grade 3
Pass required ≥24 ≤3
Omron HEM-7120 achieved
 SBP 27 2 Pass Pass
 DBP 24 2 Pass Pass
Yuwell YE680A achieved
 SBP 16 4 Fail Fail
 DBP 24 2 Pass Fail
Cofoe KF-65B achieved
 SBP 16 4 Fail Fail
 DBP 24 2 Pass Pass
Part 3 Result
Omron HEM-7120 Pass
Yuwell YE680A Fail
Cofoe KF-65B Fail