Table 2.

Protocol using the opposite arm simultaneous method based on ISO 81060-2:2018

Protocol (opposite arm simultaneous method)
Lateral difference determination (LD): 3 reference measurements on each arm
Measurements #0: reference (R0) and OptiBP (T0) measurements
 Screening and calibration for the first arm (if signal quality is satisfactory)
Measurement #1: reference (R1) and OptiBP (T1) measurements
Measurement #2: reference (R2) and OptiBP (T2) measurements
Measurement #3: reference (R3) and OptiBP (T3) measurements
 Interchange arm sides
Measurement #4: reference (R4) and OptiBP (T4) measurements
 Calibration for the second arm (if signal quality is satisfactory)
Measurement #5: reference (R5) and OptiBP (T5) measurements
Measurement #6: reference (R6) and OptiBP (T6) measurements
Measurement #7: reference (R7) and OptiBP (T7) measurements
Extra measurement #8: reference (R8) and OptiBP (T8) measurements