TABLE 2.

Detection of mycobacteria from clinical specimens in BACTEC 460 TB, MB/BacT, and LJ medium according to initial smear resultsa

Smear result (no. of isolates) No. (%) of isolates recovered in:
BACTEC 460 TB MB/BacT LJ medium
M. tuberculosis complex
 Positive (25) 23 (92)b 21 (84)b 23 (92)
 Negative (46) 46 (100) 36 (78) 25 (54)
NTM
 Positive (6) 6 (100) 5 (83) 5 (83)
 Negative (46) 37 (80) 28 (60) 13 (28)
a

A total of 3,700 specimens were tested. 

b

Loss of sensitivity due to bacterial contamination.