Skip to main content
. 2021 Nov 8;21:165. doi: 10.1186/s12880-021-00683-4

Table 1.

Comparing the measurement parameters of the NIM PET/CT phantom and NEMA IEC Body phantom

NIM PET/CT phantom NEMA IEC body phantom
CB (kBq/mL) 6.6 6.6
CH (kBq/mL) 26.4 26.4
A (kBq) 62,900 64,750
T (min) 8 8

CH, the mean activity concentration in the hot spheres; CB, the mean activity concentration in the background; A, the total activity; T, PET imaging time