Table 3.

Comparison of PET-dependent parameters between NIM PET/CT phantom and NEMA IEC Body phantom

Sphere size (mm) Contrast (%) Background variability (%) Residual error (%)
NIM PET/CT phantom 37 82.11 2.05 8.98
28 75.81 2.67
22 73.11 3.27
17 71.43 3.86
13 61.09 4.64
10 45.08 5.46
7 20.52 6.45
4
NEMA IEC Body phantom 37 82.21 1.99 9.10
28 74.84 2.53
22 72.38 3.04
17 69.01 3.59
13 58.41 4.32
10 42.52 5.33

Since the device could not detect 4 mm spheres, no measurement data of those spheres in the measurement module could be achieved. The maximum deviations are highlighted in bold