. 2021 Nov 8;21:165. doi: 10.1186/s12880-021-00683-4

Table 4.

Image quality data of the NIM PET/CT phantom on three PET/CT systems

PET/CT system Sphere size (mm) Contrast (%) Background variability (%) Residual error (%)
A 37 80.27 2.78 7.46
28 74.14 3.18
22 78.28 3.52
17 74.59 3.82
13 63.12 4.17
10 39.95 4.5
7 5.61 4.78
4
B 37 82.61 1.9 8.45
28 76.75 2.42
22 72.8 2.98
17 72.3 3.68
13 61.9 4.53
10 43.26 5.5
7 24.38 6.78
4
C 37 93.26 2.09 3.24
28 87.21 2.69
22 81.12 3.05
17 78.54 3.76
13 72.08 5.01
10 71.27 6.49
7 20.06 7.3
4 3.22 10.01

Neither A nor B system could detect 4 mm spheres, thus, data of 4 mm spheres could not be measured