Skip to main content
. 2021 Nov 2;2021:5230947. doi: 10.1155/2021/5230947

Table 3.

Usability of ECG247 versus Holter technology in patients with long-term ECG recording, n = 120 (missing data in 31 patients).

ECG247 Holter technology Diff. p
Mean usability score in different situations (SD) 1
 Showering 2.8 (3.0) 7.0 (3.7) 4.2 <0.001
 Training 1.6 (1.4) 4.6 (3.3) 3.0 <0.001
 Sleeping 1.6 (1.2) 4.2 (2.7) 2.6 <0.001
 Physical activity 1.2 (0.5) 2.9 (2.3) 1.7 <0.001
 At work 1.1 (0.5) 2.4 (2.0) 1.3 <0.001
 WC 1.0 (0.1) 2.2 (1.9) 1.2 <0.001
 Social relations 1.1 (0.3) 2.2 (2.0) 1.1 <0.001
 Eating 1.1 (0.4) 1.5 (1.1) 0.4 <0.001

Adverse events score 2
 Itching 2.8 (2.5) 3.0 (2.7) 0.2 0.49
 Erythema 2.9 (2.7) 3.1 (2.8) 0.2 0.59

System usability score3 87.4 (1.4) 67.5 (1.8) 19.9 <0.001

1Usability score: 1 (excellent)–10 (very poor). 2Adverse event score: 1 (no problem)–10 (not acceptable). 3System usability score: 0–100, >68 acceptable.