Summary of quantification results with the 3 isotopes measured on calibrated SPECT/CT images acquired on the Discovery 670 with the MEGP collimator

Isotope Real activity (MBq) Measured activity (MBq) Error (%)
223Ra 0.597 0.609 2.1
225Ac 0.401 0.386 3.5
227Th 0.404 0.356 11.8