Table 3.

Agreement between paired sensor–reference values in the range of reference glucose levels

Agreement level ± 10/10% (mmol/L) ± 15/15% (mmol/L) ± 20/20% (mmol/L) ± 30/30% (mmol/L) ± 40/40% (mmol/L)
Total 60.15 79.31 89.71 97.50 99.10
95% CI 59.30, 61.00 78.61, 80.02 89.18, 90.24 97.23, 97.77 98.94, 99.27
< 3 mmol/L 26.92 34.62 42.31 46.15 50.00
95% CI 9.87, 43.97 16.33, 52.90 23.32, 61.30 26.99, 65.32 30.78, 69.22
3–3.9 mmol/L 42.67 52.00 66.67 77.33 86.67
95% CI 31.47, 53.86 40.69, 63.31 56.00, 77.34 67.86, 86.81 78.97, 94.36
3.9–10 59.60 78.72 89.08 97.62 99.31
95% CI 58.60, 60.61 77.89, 79.56 88.45, 89.72 97.30, 97.93 99.15, 99.48
10–13.9 61.97 81.35 92.42 98.59 99.45
95% CI 60.14, 63.80 79.89, 82.82 91.42, 93.41 98.15, 99.04 99.17, 99.73
> 13.9 63.45 83.74 91.65 95.83 98.13
95% CI 59.87, 67.03 81.00, 86.48 89.60, 93.71 94.34, 97.31 97.12, 99.14