Table 3.
Agreement between paired sensor–reference values in the range of reference glucose levels
| Agreement level | ± 10/10% (mmol/L) | ± 15/15% (mmol/L) | ± 20/20% (mmol/L) | ± 30/30% (mmol/L) | ± 40/40% (mmol/L) |
|---|---|---|---|---|---|
| Total | 60.15 | 79.31 | 89.71 | 97.50 | 99.10 |
| 95% CI | 59.30, 61.00 | 78.61, 80.02 | 89.18, 90.24 | 97.23, 97.77 | 98.94, 99.27 |
| < 3 mmol/L | 26.92 | 34.62 | 42.31 | 46.15 | 50.00 |
| 95% CI | 9.87, 43.97 | 16.33, 52.90 | 23.32, 61.30 | 26.99, 65.32 | 30.78, 69.22 |
| 3–3.9 mmol/L | 42.67 | 52.00 | 66.67 | 77.33 | 86.67 |
| 95% CI | 31.47, 53.86 | 40.69, 63.31 | 56.00, 77.34 | 67.86, 86.81 | 78.97, 94.36 |
| 3.9–10 | 59.60 | 78.72 | 89.08 | 97.62 | 99.31 |
| 95% CI | 58.60, 60.61 | 77.89, 79.56 | 88.45, 89.72 | 97.30, 97.93 | 99.15, 99.48 |
| 10–13.9 | 61.97 | 81.35 | 92.42 | 98.59 | 99.45 |
| 95% CI | 60.14, 63.80 | 79.89, 82.82 | 91.42, 93.41 | 98.15, 99.04 | 99.17, 99.73 |
| > 13.9 | 63.45 | 83.74 | 91.65 | 95.83 | 98.13 |
| 95% CI | 59.87, 67.03 | 81.00, 86.48 | 89.60, 93.71 | 94.34, 97.31 | 97.12, 99.14 |