Table 6.

Stability evaluation of CGMS at different validation stages

Period of CGMS wearing Visit CGMS venous blood glucose matching number Consistency rate of reference value 10/10% Consistency rate of reference value 15/15% Consistency rate of reference value 20/20% Consistency rate of reference value 30/30% Consistency rate of reference value 40/40% MARD (%)
Initial stage Day 1 or 2 4284 2410 (56.26%) 3341 (77.99%) 3857 (90.03%) 4197 (97.97%) 4240 (98.97%) 10.43
Intermediate stage Day 7 ± 1 4260 2737 (64.25%) 3526 (82.77%) 3923 (92.09%) 4196 (98.50%) 4255 (99.88%) 9.12
Final stage Day 14 4144 2485 (59.97%) 3196 (77.12%) 3602 (86.92%) 3978 (95.99%) 4079 (98.43%) 10.78

CGMS continuous glucose monitoring system, MAR mean absolute relative error