Table 6.
Stability evaluation of CGMS at different validation stages
Period of CGMS wearing | Visit | CGMS venous blood glucose matching number | Consistency rate of reference value 10/10% | Consistency rate of reference value 15/15% | Consistency rate of reference value 20/20% | Consistency rate of reference value 30/30% | Consistency rate of reference value 40/40% | MARD (%) |
---|---|---|---|---|---|---|---|---|
Initial stage | Day 1 or 2 | 4284 | 2410 (56.26%) | 3341 (77.99%) | 3857 (90.03%) | 4197 (97.97%) | 4240 (98.97%) | 10.43 |
Intermediate stage | Day 7 ± 1 | 4260 | 2737 (64.25%) | 3526 (82.77%) | 3923 (92.09%) | 4196 (98.50%) | 4255 (99.88%) | 9.12 |
Final stage | Day 14 | 4144 | 2485 (59.97%) | 3196 (77.12%) | 3602 (86.92%) | 3978 (95.99%) | 4079 (98.43%) | 10.78 |
CGMS continuous glucose monitoring system, MAR mean absolute relative error