Table 3.
Type 2 phantoms measured with device id50. See text for details. Values reported for each phantom are average and standard deviation (and corresponding CV) for the three different days of measurement.
| Phantom | TSI | THC | ||
|---|---|---|---|---|
| (%) | (%) | ( M) | (%) | |
| #1 | 50.6 ± 0.8 | 1.6 | 42.4 ± 0.4 | 0.9 |
| #2 | 50.2 ± 0.4 | 0.9 | 44.1 ± 0.4 | 1.0 |
| #3 | 50.6 ± 0.7 | 1.4 | 44.5 ± 0.6 | 1.3 |
| #4 | 51.7 ± 0.9 | 1.6 | 41.6 ± 1.0 | 2.4 |
| #5 | 51.5 ± 0.6 | 1.3 | 41.3 ± 0.7 | 1.7 |
| #6 | 52.7 ± 0.5 | 0.9 | 39.9 ± 0.1 | 0.3 |
| #7 | 51.2 ± 0.4 | 0.7 | 42.5 ± 0.3 | 0.7 |
| #8 | 54.5 ± 0.4 | 0.8 | 37.2 ± 0.8 | 2.2 |
| #9 | 50.4 ± 0.6 | 1.3 | 44.6 ± 0.7 | 1.5 |
| #10 | 54.2 ± 0.7 | 1.2 | 40.9 ± 1.0 | 2.5 |