. 2021 Jun 9;31(12):9120–9130. doi: 10.1007/s00330-021-08072-9

Table 3.

Intraclass correlation coefficients (ICCs), standard error of measurement (SEM), and minimum detectable difference (MDD) values for interreader reliability and test-retest reliability

Interreader ICC Test-retest ICC SEM MDD
Reader 1 Reader 2 Interreader Test-retest Interreader Test-retest
MTR 0.81 (0.60–0.92) 0.79 (0.62–0.90) 0.75 (0.55–0.88) 1.7% 1.7% 4.7% 4.7%
T2 0.92 (0.81–0.97) 0.92 (0.84–0.96) 0.9 (0.80–0.95) 2.67 ms 2.66 ms 7.4 ms 7.37 ms
PSD 0.94 (0.85–0.98) 0.94 (0.87–0.97) 0.94 (0.88–0.97) 21.3 20.1 59 55.7
Normalized PSD 0.89 (0.74–0.95) 0.9 (0.81–0.96) 0.89 (0.78–0.95) 0.03 0.03 0.08 0.08

Intraclass correlation coefficients (ICCs) with 95% confidence-interval were calculated according to Shrout and Fleiss. SEM and MDD calculations were based on Popovic and Thomas [31]. MTR magnetization transfer ratio, T2 transverse relaxation time, PSD proton spin density