TABLE III.

Comparison of the ratio ϒ of the volume of the homogeneous field region to that of the device itself from various designs of the magnetostatic cavities. The dimensions are a diameter Ø and length l for the mu-metal assembly of cylinder body and end caps for a MSS, and a width w, height h and a length l for a magic box. dh and lh are the diameter and length of the cylinder of the homogeneous field region. The data of the magnetostatic cavity devices of the last 6 rows are from other labs.

cavity
name
cavity
type
dimensions
cm × cm
ϒ instrument
class
P or A dh × lh
cm × cm
BB
10−4 cm−1
Gemini solenoid 25.4 × 35.6 0.064 TAS/SANS26 analyzer 12 × 10 2.8 ± 0.1
Kronos solenoid 20.3 × 25.4 0.14 TAS/SANS25 polarizer 12 × 10 6.0
Venus solenoid 25.4 × 33 0.061 SANS analyzer 12 × 9 4.4 ± 0.2
Honesty solenoid 27.9 × 35.6 0.055 Reflectometer27 analyzer 13 × 9 3.7 ± 0.2
Nyx solenoid 25.1 × 29.2 0.078 TAS polarizer 12 × 10 4.2 ± 0.2
Flattop magic box 40 × 15 × 28.4 0.066 TAS17 polarizer 12 × 10 5.7 ± 0.4
solenoid24 27.2 × 36.4 0.037 polarizer 10 × 10 <4.0
solenoid 30 × 46 0.024 TAS21 analyzer 10 × 10 2.0
solenoid 9.5 × 20 0.030 SCD10 polarizer 3 × 6 45.0
solenoid 20 × 30 0.030 POLANO23 polarizer 6 × 10 <10.0
magic box 20 × 40 × 78 0.015 TOPAS35 polarizer 12 × 8 4.4
magic box 30 × 40 × 62 0.012 MARIA35 analyzer 12 × 8 <2.0
magic box18 40 × 17× 80 0.005 6 × 10 <2.0
solenoid19 33 × 96 0.003 8 × 8 <1.0