Summarized results of PCM and the direct “on chip quantification” demonstrating the FS 100% bacteria capture efficiency.

Biol Rep 1 Biol Rep 2 Biol Rep 3
Tech Rep 1
[Ave cfu/mL]
Tech Rep 2
[Ave cfu/mL]
Tech Rep 3
[Ave cfu/mL]
Tech Rep 1
[Ave cfu/mL]
Tech Rep 2
[Ave cfu/mL]
Tech Rep 3
[Ave cfu/mL]
Tech Rep 1
[Ave cfu/mL]
Tech Rep 2
[Ave cfu/mL]
Tech Rep 3
[Ave cfu/mL]
Plate-count method [cfu/mL] “PCM quantification”
NegCTRL (FS control buffer) 0 0 0 0 0 0 0 0 0
Influent ≤ 250 cfu/mL 70 20 17 52 12 42 65 53 20
Effluent 0 0 0 0 0 0 0 0 0
Capeff [%] 100 100 100 100 100 100 100 100 100
Direct FS baseline of particles “on-chip quantification” [particles/mL]
FS baseline of particles 6 15 3 4 3 12 26 13 5
Direct total particle capture “on-chip quantification” [particles/mL]
Total particle capture 184 130 198 124 179 189 124 128 139
Direct “on-chip quantification” (Total particle capture on-chip minus FS baseline of particles on-chip) [bacteria/mL]
Total capture 178 115 195 120 176 177 98 115 134

Capeff, Capture efficiency. The flow rate used in the Fluid-Screen bacterial capture and separation experiments is 300 µL/min.