Table 1.

Parameters for the 2D and 3D MRE

Parameters 2D MRE 3D MRE
Acquisition matrix 80 × 80 80 × 80
TE (ms) 51.4–52.5 51.7–52.8
TR (ms) 1000–1067 1334
EPI-shots 1 1
No. of breath holds × breath hold time (s) 1 × 16 3 × 21
FOV (cm) 38–44 38.7–44.8
No. of slices × slice thickness (mm) 4 × 8 32 × 3.5
Gap between slices (mm) 2 0
Parallel imaging acceleration factor 2 2
Phase offsets 4 3
Receiver bandwidth 500 kHz 500 kHz
% FOV in phase-encoding direction 75–100 100
Flow compensation Yes Yes
Superior-inferior spatial presaturation bands Yes Yes
Motion sensitivity (μm/radian) 7.86 7.86
Motion encoding directions ± Z ± X, Y, Z

TE, echo time; TR, repetition time; FOV, field of view; 2D, two-dimensional; 3D, three-dimensional; EPI, echo-planar imaging; MRE, magnetic resonance elastography