Hemolysis and platelet aggregation upon blood contact with polymer grafts (%) (Me [25; 75])

Graft type Degree of hemolysis Maximum platelet aggregation
Intact platelet-rich plasma 14.61 [13.63; 17.72]
PHBV/PCL/GFmix 0 [0; 0] 17.25 [16.30; 17.96]
PHBV/PCL/GFmix + PVP + heparin 0.36 [0.36; 0.72]* 26.78 [25.47; 31.53]*+
PHBV/PCL/GFmix + PVP + heparin + iloprost 0.36 [0.36; 0.36]* 8.22 [8.13; 8.78]*+