Analysis of performance parameters per method in biomarker dataset (left) and 94-replicates-4-dilutions set (right). For each method, the mean rank is given for each of the performance indicators bias, RE, CV, precision and resolution. The methods are sorted based on the average of these ranks

Methods Bias(10,000/0.001) Bias(1) Relative error Coefficient of variation Precision Resolution rank
CqMAN 2/2 1/1 3/1 1/2 1/1 4/2 2.08/1.58
Cy0 1/1 1/1 1/2 2/1 4/2 3/1 2.08/1.92
LinRegPCR 3/6 2/4 4/4 4/3 3/3 1/4 3.00/3.67
PCR-Miner 7/3 3/2 2/3 3/4 2/4 2/3 3.33/3.33
5PSM 6/8 7/7 8/8 5/7 5/7 6/5 6.33/7.67
DART 5/7 6/6 7/7 6/6 6/6 5/6 6.00/6.67
FPLM 4/5 5/3 6/6 7/8 7/8 7/7 6.17/6.17
FPK-PCR 8/4 4/5 5/5 8/5 8/5 8/8 7.00/5.00