Design constraints and specifications for the arborizing catheter

Design criteria Specification
Max cannula OD 3 mm
Length (for animal model) 13 cm
Max microneedle OD × ID 500 μm × 200 μm
Max microneedle deployment depth 5 cm
Number of ports ≥5
Microneedle separation distance >1–1.5 cm
Materials Ceramic, titanium, PEEK