Skip to main content
. 2021 Jun 11;34(11):doab036. doi: 10.1093/dote/doab036

Table 2.

Reason for device removal and mean time to removal

Reason for removal Number of removals Percentage of total removals Mean time to removal, months (±SD)
Dysphagia 292 47.9 10.9 (11.9)
Persistent GERD 125 20.5 20.5 (13.0)
Erosion 27 4.4 25.0 (12.9)
Abdominal pain/pain 46 7.6 15.8 (14.3)
Discontinuous device 17 2.8 33.7 (6.0)
Need for MRI 11 1.8 28.6 (13.2)
Vomiting 16 2.6 7.4 (8.2)
Gastroparesis 4 0.7 20.7 (18.5)
Device migration 3 0.5 12.6 (17.7)
Other/unknown 68 11.2 6.8 (6.4)
Total removals 609 100.0 14.6 (13.4)

Discontinuous devices were the result of a manufacturing issue that resulted in a voluntary recall in 2018.