Table 3.
TR-FRET reading settings using TECAN F500 plate reader
Reading settings in Tecan F500 plate reader | |
---|---|
Fluorescence 1: d2 | |
Excitation wavelength | 340 nm |
Emission wavelength | 665 nm |
Excitation bandwidth | 20 nm |
Emission bandwidth | 8 nm |
Gain | Optimal |
Number of flashes | 10 |
Integration time | 500 μs |
Lag time | 150 μs |
Fluorescence 2: Tb | |
Excitation wavelength | 340 nm |
Emission wavelength | 620 nm |
Excitation bandwidth | 20 nm |
Emission bandwidth | 10 nm |
Gain | Optimal∗ |
Number of flashes | 10 |
Integration time | 500 μs |
Lag time | 150 μs |
When reading TR-FRET multiple times (every hour or in a kinetics mode, for example) the Gain of the donor signal should be set to a fixed value corresponding to the value obtained in the first reading.