Table 3.
Measurement Floor | Mean MD (dB) at Measurement Floor | Absolute Dynamic Range | Relative Dynamic Range (%) | Steps to the Measurement Floor | CRw | |
---|---|---|---|---|---|---|
OCT Metrics | ||||||
RNFL, µm | 60.1 | −17.8 ± 0.927 | 28.1 | 31.8 | 7.20 | 3.90b |
GCIPL, µm | 54.0a | −23.6 ± 1.14 | 23.5 | 30.4 | 7.33 | 3.21c |
Peripapillary OCTA Metrics | ||||||
Custom | ||||||
pVAD | 0.211 | −17.6 ± 1.06 | 0.139 | 39.8 | 2.14 | 0.065d |
Commercial | ||||||
pPDZ | 0.322a | −26.6 ± 1.53 | 0.102 | 24.1 | 4.25 | 0.024d |
Macular | ||||||
OCTA Metrics | ||||||
Custom | ||||||
mVAD | 0.322a | −21.8 ± 1.03 | 0.151 | 31.9 | 3.87 | 0.039d |
mVSD | 0.118a | −22.0 ± 1.03 | 0.069 | 36.7 | 3.81 | 0.018d |
Commercial | ||||||
mPDZ | 0.358a | −24.9 ± 1.03 | 0.068 | 15.9 | 1.26 | 0.054d |
mVDZ | 14.8a | −24.2 ± 1.03 | 2.94 | 16.5 | 1.37 | 2.14d |
Measurement floors were estimated for these metrics using an arbitrary change-point MD of −20 dB per protocol delineated in text. Mean MD across measurement floors are reported with standard errors.
Derived from Mwanza et al. (2010)30.
Derived from pooled test-retest standard deviation data from Mwanza et al. (2011)31.
These CRw values were derived from repeatability data sets comprising non-glaucoma subjects at the USC Roski Eye Institute15.
MD = mean deviation, CRw = coefficient of repeatability, OCT = optical coherence tomography, RNFL = retinal nerve fiber layer, µm = micrometers, GCIPL = ganglion cell-inner plexiform layer, OCTA = optical coherence tomography angiography, pVAD = peripapillary vessel area density, pPDZ = peripapillary perfusion density, mVAD = macular vessel area density, mVSD = macular vessel skeleton density, mPDZ = macular perfusion density, mVDZ = macular vessel density.