Skip to main content
. 2021 Nov 24;300:114378. doi: 10.1016/j.jviromet.2021.114378

Table 1.

Detection limits using the SARS-CoV-2 Analytical Qnostic panel 01. The limit of detection for the SARS-CoV-2 in-house qPCR and the N-gene using the GeneFinder was 50 dC/mL. The RdRp-gene and E-gene using the GeneFinder was 500 dC/mL.

SARS-CoV-2 Qnostic analytical panel
In-house SARS-CoV-2 GeneFinderTM COVID-19 Plus RealAmp kit
Panel number dC/mL Log10 dC/mL E- gene (Ct value) RdRP- gene (Ct value) N- gene (Ct value) E-gene (Ct value)
SCVA2AQP01-S01 1000000 6,00 21.87 24.22 24.17 22.27
SCVA2AQP01-S02 100000 5,00 24.66 27.09 27.11 25.46
SCVA2AQP01-S03 10000 4,00 27.87 30.07 29.74 28.77
SCVA2AQP01-S04 5000 3.7 28.52 31.11 30.62 29.73
SCVA2AQP01-S05 1000 3,00 30.57 34.39 32.8 32.88
SCVA2AQP01-S06 500 2.7 31.89 43.97 34.08 35.45
SCVA2AQP01-S07 100 2,00 34.26 Not detected 36.4 Not detected
SCVA2AQP01-S08 50 1.7 35.01 Not detected 40.14 Not detected
SCVA2AQP01-S09 Negative Not detected Not detected Not detected Not detected