Intra- and inter- assay variation of the GeneFinder ™ COVID-19 Plus RealAmp kit using the ELITe InGenius platform. The intra- assay variation of a sample with a high viral load (Ct around 20) was lower compared to a sample with a low viral load (Ct around 31). In contrast, the inter assay variation was comparable for the high and low viral load samples.

Gene Sample type Intra-assay Inter-assay
Average Ct ± SD (%CV) Average Ct ± SD (%CV)
E-gene high positive 19.6 ± 0.48 (2.45) 18.97 ± 1.12 (5.92)
low positive 30.93 ± 1.44 (4.66) 31.32 ± 0.98 (3.11)
RdRp-gene high positive 22.17 ± 0.29 (1.31) 22.44 ± 0.89 (3.95)
low positive 34.55 ± 1.22 (3.54) 34.15 ± 1.57 (4.58)
N-gene high positive 20.25 ± 0.15 (0.72) 20.20 ± 1.02 (5.07)
low positive 31.09 ± 1.63 (5.23) 31.87 ± 0.82 (2.59)