TABLE 5.
Examples of Errors Considered to Be Preventable With Smart Infusion Device
| Preventability With a Smart Infusion Device | Example |
|---|---|
| Not preventable | Actrapid (soluble insulin) infusion expired but still running (observational data; category C) |
| Preventable with any pump | Vancomycin 2 g in 250 mL running faster than recommended rates through gravity feed (observational data; category E) |
| Possibly preventable with a standalone smart infusion device depending on limits set in drug library | Noradrenaline standard prescription chart specified 4 mg in 50 mL; 8 mg in 50 mL administered (observational data; category D) |
| Possibly preventable with a smart infusion device integrated with an electronic prescribing and barcode administration system | Piperacillin/tazobactam prescription specified a 3-h infusion due to impaired renal function but the nurses gave it via a bolus dose, which is what they would normally do (observational data; category D) |