Table 1.
Data of MTF (50 lp/mm) and Strehl (mean) of the 2 IOLs
| 3 mm centered | MTF mean (SG lp/mm) | Strehl mean | 4.S mm centered | MTF mean (SG lp/mm) | Strehl mean |
|
| |||||
| Vivity | G.2SG | G.261 | Vivity | G.2G2 | G.176 |
| LuxSmart | G.2S7 | 0.355 | LuxSmart | 0.24З | 0.206 |
|
| |||||
| З mm decentered | MTF mean (SG lp/mm) | Strehl mean | 4.S mm decentered | MTF mean (SG lp/mm) | Strehl mean |
|
| |||||
| Vivity | 0.266 | G.272 | Vivity | 0.126 | 0.133 |
| LuxSmart | G.247 | 0.234 | LuxSmart | G.21S | 0.183 |
| З mm tilt | MTF mean (SG lp/mm) | StrehI mean | 4.S mm tilt | MTF mean (SG lp/mm) | Strehl mean |
| Vivity | G.221 | 0.232 | Vivity | G.214 | G.22S |
| LuxSmart | 0.360 | G.428 | LuxSmart | G.229 | G.211 |
MTF, modulation transfer function; IOL, intraocular lens.