Skip to main content
. 2021 Oct 22;10(11):1661. doi: 10.3390/antiox10111661

Table 3.

Descriptions of the analytical methods previously developed to analyze MDA in EBC with DNPH derivatization.

Reference Technique EBC
Volume (µL)
LOD (pg/mL) LOQ (pg/mL) Intermediate Precision (%) Recovery (%) N a EBC
Levels (pg/mL)
This study LC-ESI-MS/MS 125 70 211 9 92 164 115 (<LOD-886) b
[29] LC-ESI-MS 100 42 126 12 92 205 491 (294–711) c
[28] LC-APCI-MS/MS 100 72 216 7 n.a. 12 806 (77–1685) b

a: number of samples analyzed; b: median with the range in parentheses; c: median above LOD with interquartile range into parentheses.