Table 1.

Ventilator settings and limits.

Conventional Ventilation INTELLiVENT-ASV
Ventilator Settings
etCO2 target 4.5 to 6.0 kPa 4.5 to 5.5 kPa
SpO2 target 90 to 92% 89 to 93%
Tidal volume target 5 to 8 mL/kg PBW 5 to 8 mL/kg PBW
Ventilator Limits
Maximum airway pressure 30 cm H2O 30 cm H2O
PEEP limit ≤12 cmH2O 5 to 12 cmH2O
Tidal volume alarm limit 9 mL/kg PBW 9 mL/kg PBW
Respiratory rate limit 30 per min 30 per min
FiO2 upper limit 0.60 0.60

Abbreviations: etCO2, end-tidal carbon dioxide; SpO2, peripheral oxygen saturation; PEEP, positive end expiratory pressure; FiO2, fraction of inspired oxygen. With INTELLiVENT-ASV, the target ranges for etCO2 and SpO2 are pre-specified and set automatically; when a certain target is chosen, the lower and upper limits of the target ranges are 0.5 kPa under and above the target of 5 kPa for etCO2, and 2% around the 91% for SpO2.