Table 4.
ID Now (date) | C2R (min) | Reference | Assay/instrument | Ct values / RLU | Resolution |
---|---|---|---|---|---|
NEG (03/20) | 97 | POS (03/20) | Allplex/MGI | 31.85/34.69/37.05 | FN ID Now |
NEG (04/22) | 43 | POS (04/22) | Allplex/MGI | 25.27/28.69/26.71 | FN ID Now |
NEG (04/23) | 48 | POS (04/23) | Aptima/Hologic | RLU: 1279 | FN ID Now |
POS (05/02) | - | NEG (05/02) | Allplex/MGI and cobas® 6800 | Not detected | FP ID Now |
C2R = time from collection of the ID Now sample to result reporting; FN = false negative; TP = true positive; FP = false positive; RLU = relative light units.