Summary of the recommended technical parameters and accepted alternatives for WBLDCT in MM [17]

Parameters Recommendation Alternative
No. of detector rows At least 16
Field of view From the cranial vertex to the prox. metaphyses of both tibiae incl. both humeri
Regular patient positioning Supine with both arms aside and above the skull with as little bending of the elbows as possible Supine with both arms alongside the body underlaid with armrests to lift them above spine level
Positioning of obese or PRM Supine with both arms placed in front of the body with both hands folded
Tube voltage (kV) 120 140
Tube current–time product (mAs) 50–70 14–25
Collimation 0.5–1.5 mm
Reconstruction convolution kernel Combination of a sharp, high-frequency “bone” kernel and a smoother “soft-tissue” kernel One singular intermediate, middle-frequency kernel permitting acquisition of all images with different window settings*
Iterative reconstruction algorithms Requested, if possible, to reduce image noise and artifacts while reducing radiation exposure
Thickness/increment of axial slices 2/1 mm 3/1.5 mm
MPRs requested in sagittal and coronal direction and parallel to the long axis of the prox. extremities

Abbreviations: No.: number; prox.: proximal; incl.: inclusive; PRM: persons with reduced mobility; kV: kilovoltage; mAs: milliampere-seconds; mm: millimeters; MPRs: multiplanar reconstructions

* With the limitation that attenuation measurements of peripheral intramedullary lesions should not be performed on bone kernel-reconstructed images that were set on the soft-tissue window, but on soft-tissue kernel-reconstructed images