Published in final edited form as: Med Phys. 2020 Jan 6;47(3):958–974. doi: 10.1002/mp.13983

Table II.

Summary of three-dimensional (3D) imaging protocols for the mobile C-arm in this work. Clinical protocols are in white, and custom protocols are in gray. The mAs/view and mAs total values are set by automatic exposure control (AEC) (with values shown for 16 and 32 cm diameter head and body acrylic cylindrical phantoms).

Protocols Tube voltage (kV) Number of projections Scan time (s) Angular step (Δθ) Pixel binning Tube output (mAs/view) Tube output (total) (mAs)
16 cm cylindrical phantom (head)
 Low dose (L) 110 100 30 2 × 2 0.05 5
 Standard (S) 110 200 30 2 × 2 0.05 10
 High quality (H) 110 400 30 0.5° 2 × 2 0.08 32
 Extra power obese (O) 110 400 60 0.5° 2 × 2 0.08 32
 Custom head 1 110 100 30 2 × 2 0.57 57
 Custom head 2 110 200 30 2 × 2 0.57 114
 Custom head 3 110 400 30 0.5° 2 × 2 0.68 272
32 cm cylindrical phantom (body)
 Custom body 1 110 100 30 2 × 2 0.16 16
 Custom body 2 110 200 30 2 × 2 0.16 32
 Custom body 3 110 400 30 0.5° 2 × 2 0.22 88
 Low dose (L) 110 100 30 2 × 2 1.01 101
 Standard (S) 110 200 30 2 × 2 1.02 204
 High quality (H) 110 400 30 0.5° 2 × 2 1.47 588
 Extra power obese (O) 110 400 60 0.5° 2 × 2 1.65 660