Summary of dose measurements for clinical and custom three-dimensional (3D) imaging protocols on the mobile C-arm for head and body protocols. Custom protocols are highlighted in gray. The terms D0, D¯P, and DW represent the central, average peripheral, and weighted dose, respectively, as defined in Section 2.B.2

Protocol D0 (mGy) D¯P (mGy) Dw (mGy)
16 cm cylindrical phantom (head)
 Low dose 0.7 0.8 0.8
 Standard 1.4 1.6 1.5
 High quality 4.2 4.9 4.6
 Extra power: obese 4.2 4.9 4.7
 Custom head 1 7.6 8.7 8.3
 Custom head 2 15.1 17.4 16.6
 Custom head 3 36.0 41.6 39.7
32 cm cylindrical phantom (body)
 Custom body 1 0.6 1.1 0.9
 Custom body 2 1.2 2.1 1.8
 Custom body 3 3.2 5.8 4.9
 Low dose 4.4 7.2 6.3
 Standard 8.9 14.4 12.5
 High quality 20.0 37.1 31.4
 Extra power: obese 25.1 44.5 38.0