Table 2.

The means of the test object reading results for each DR system averaged across the exposure settings.

Test objects BodiiRay Fujifilm FDR Go Siemens YSIO Max
Line pair (lp/mm) 3.033 3.033 2.900
Contrast detail depth (inch) 0.029 0.036 0.030
Contrast detail diameter (inch) 0.091 0.093 0.094
Step wedge (step) 11.000 10.222 11.000