. 2000 Mar;38(3):1203–1208. doi: 10.1128/jcm.38.3.1203-1208.2000

TABLE 1.

MODS sensitivity and time to detection of MTB versus standard culture methods

Method No. of samples positive by method No. of samples positive by at least one culturea Sensitivity (%)b Median detection time (range)c
Auramine O stain 76 98 78
MODS 89 97 92 9 (4–31)
MGIT 88 95 93 10 (4–39)
LJ 73 96 76 24 (6–59)
microagar 7H11 75 96 78 14.5 (4–28)
PCR 81 90 90
a

Only samples which were positive by at least one culture (MODS, MGIT, LJ, or 7H11) and interpretable by the given method were considered in determining sensitivity. Of the 172 samples, 98 were positive by at least one culture. Denominators differ for different methods because not all results were interpretable due to contamination or (in the case of PCR) insufficient patient sample. 

b

Sensitivity is based on the gold standard of detection by any culture. 

c

Includes only interpretable results. Values shown are in days.