Skip to main content
. 2000 Mar;38(3):1247–1249. doi: 10.1128/jcm.38.3.1247-1249.2000

TABLE 2.

Mean log10 differences of RNA values for each HIV-1 subtype with respect to the Roche Amplicor Monitor 1.5 test

HIV subtype (n)a Mean log10 differenceb
Roche 1.0 Gen-Probe Bayer 3.0 Digene 1.0 NucliSens 2.0
A (3) 2.58 0.02 0.41 0.32 0.29
B (7) 0.34 0.12 0.09 0.12 0.09
C (5) 0.15 0.07 0.23 0.19 0.06
D (3) 0.13 −0.11 0.11 0.14 −0.19
E (8) 1.18 −0.15 0.16 0.30 0.49
F (3) 1.82 −0.02 0.40 0.35 0.92
G (1) 0.47 −0.36 0.29 0.18 NAc
Mean difference (30) 0.89 −0.02 0.20 0.23 0.27
SD 0.87 0.28 0.18 0.18 0.35
a

Number of HIV isolates in each subtype category. 

b

Mean log10 differences of RNA values for each HIV-1 subtype by HIV-1 RNA quantitation tests compared to the values found by the Roche Amplicor Monitor version 1.5 test. The HIV-1 RNA quantitation tests used were the Roche Amplicor HIV-1 Monitor test (version 1.0), Bayer Quantiplex HIV RNA (version 3.0), Digene Hybrid Capture II HIV RNA test (version 1.0), Organon Teknika NucliSens HIV QT test (version 2.0), and Gen-Probe HIV-1 Viral Load Assay (in development). 

c

NA, not applicable.