TABLE 2.
HIV subtype (n)a | Mean log10 differenceb
|
||||
---|---|---|---|---|---|
Roche 1.0 | Gen-Probe | Bayer 3.0 | Digene 1.0 | NucliSens 2.0 | |
A (3) | 2.58 | 0.02 | 0.41 | 0.32 | 0.29 |
B (7) | 0.34 | 0.12 | 0.09 | 0.12 | 0.09 |
C (5) | 0.15 | 0.07 | 0.23 | 0.19 | 0.06 |
D (3) | 0.13 | −0.11 | 0.11 | 0.14 | −0.19 |
E (8) | 1.18 | −0.15 | 0.16 | 0.30 | 0.49 |
F (3) | 1.82 | −0.02 | 0.40 | 0.35 | 0.92 |
G (1) | 0.47 | −0.36 | 0.29 | 0.18 | NAc |
Mean difference (30) | 0.89 | −0.02 | 0.20 | 0.23 | 0.27 |
SD | 0.87 | 0.28 | 0.18 | 0.18 | 0.35 |
Number of HIV isolates in each subtype category.
Mean log10 differences of RNA values for each HIV-1 subtype by HIV-1 RNA quantitation tests compared to the values found by the Roche Amplicor Monitor version 1.5 test. The HIV-1 RNA quantitation tests used were the Roche Amplicor HIV-1 Monitor test (version 1.0), Bayer Quantiplex HIV RNA (version 3.0), Digene Hybrid Capture II HIV RNA test (version 1.0), Organon Teknika NucliSens HIV QT test (version 2.0), and Gen-Probe HIV-1 Viral Load Assay (in development).
NA, not applicable.